MOONEY VISCOMETER
MOONEY VISCOMETER (MV Xgen 100)
Specifications
Mooney Measuring Range:
|
0 to 200 MU (4 digits)
|
Pre-heat time:
|
0 to 10 minutes
|
Mooney Measurements:
|
Choice of 3 readings
|
Scorch points:
|
Choice of 4 readings
|
Rotor Size:
|
Large and small
|
Rotor speed:
|
2.0 +- 0.02 RPM
|
Drive Motor:
|
Synchronous geared motor
|
Temperature:
|
PT100 Platinum resistors
Microprocessor controlled. Calibrated range 50 to 200 degree celsius Independent Upper and Lower Platen Control. |
Temperature Control:
|
Accuracy +- 0.5 degree celsius
|
Recording & Display:
|
Computer controlled testing.
Directly on-line display. Memory storage of data. Automatic computation of results. Multiple display of graphs. Statistical Analysis. Statistical Quality Control. |
Printed Data:
|
A4
|
Electric Power Supply:
|
AC 175 – 275V, 50 Hz, 20 Amp. Max.
|
Compressed air supply:
|
60 psi (4.2 kg / sq.cm) minimum operating pressure controlled by external regulator with gauge.
|
Environment:
|
Dust free reasonably controlled ambient temperature and humidity.
|
Dimensions:
|
Main panel – 60 x 26 x 23 inches
Computer System – Table Top. |